Sharpness and Noise


Description

Sharpness and Noise window

Measures the resolution and the noise of selected areas of the current slice, detects astigmatism and gives focus scores. Every calculation adds rows to the results table, so different images, slices or microscope settings can be compared side by side. The table is kept until MIB is closed.

The analysed area should hold even, unstructured texture: a gold-on-carbon test specimen, cytoplasm or embedding resin. Membranes, fibres, edges and other oriented structures are read as astigmatism. Use at least 256 x 256 pixels for stable values.

Methods

Methods panel

Tick the methods to calculate; each fills its own columns of the results table.

  • FFT resolution and astigmatism - from the power spectrum of the area1: the finest detail that still stands out from the noise, overall and in each direction, and the noise of the image.
    • FFT res. - the resolution.
    • Best res., Worst res. - the resolution in the sharpest and in the most blurred direction.
    • Astig. ratio - worst divided by best resolution; 1 means no astigmatism.
    • Astig. angle - the most blurred direction, counterclockwise from horizontal; it is the direction of the streaks seen in an astigmatic image. With astigmatism it turns by about 90 degrees when the focus passes from under- to overfocus.
    • Noise SD - the noise in grey values.
    • SNR, dB - the signal-to-noise ratio of the whole area, signal power over noise power, in decibels: 0 dB means equal power, 10 dB ten times, 20 dB a hundred times more signal than noise.
  • Decorrelation resolution - an independent estimate of the resolution2 (Decorr. res.), less affected by noise, without directions. When it differs a lot from the FFT value, or gives no value, the area does not suit the method; rely on the FFT value.
  • Sharpness scores - two focus scores without units3 (Grad. energy, Lapl. var.), higher is sharper. Use them only to rank images of the same specimen, for example a focus series.

Region

Region panel

The dropdown selects which part of the current slice is analysed, and the button next to it starts the analysis: Calculate, or Draw for Drawn rectangle. Every analysis adds its results to the table; a long one can be stopped with Cancel, and nothing is added then. All modes use the image at its full resolution, whatever the zoom of the image view.

  • Drawn rectangle - a rectangle drawn on the image. Press Draw, drag over the image, adjust the rectangle if needed and double-click inside it; the area is analysed straight away (Esc cancels). The next press of Draw shows the previous rectangle again when it is in view: double-click it to analyse the same area again, for example on another slice or with other settings, or move and resize it first. To cancel at that point, delete the rectangle from its right-click menu. After switching to another dataset or orientation a new rectangle is drawn.
  • Mask layer - every separate object of the mask on the current slice gives its own row; the rectangle enclosing each object is analysed. Useful to keep the analysed areas while the selection layer is used for other work.
  • ROIs - every ROI of the current orientation is analysed and gives its own row, named after the ROI label. ROIs are defined in the ROI panel; the rectangle enclosing each ROI is analysed. Use several ROIs across the field of view to see whether the focus or the astigmatism changes from the centre to the edges.
  • Selection layer - the same as Mask layer, for the objects of the selection layer. A quick way to mark several areas without creating ROIs: draw them with any selection tool. Objects smaller than 64 x 64 pixels are skipped.
  • Shown area - the part of the slice visible in the image view. Zoom and pan to the area of interest, then press Calculate. The quickest way to check a spot.
  • Whole slice - the complete slice. A large slice takes longer, and usually mixes structures of different orientation, so prefer a smaller area for astigmatism.

Color channel selects the channel to analyse in multichannel images.

Settings

Settings panel

  • SNR threshold - decides where the FFT method puts the resolution limit. Fine details of an image correspond to high spatial frequencies, coarse structures to low ones. The method works in three steps:

    1. It averages the power of the image at each spatial frequency; this is the blue curve of the Radial profile tab.
    2. It measures the noise floor: the power of the noise alone, at the highest frequencies, where an image holds no signal.
    3. Going from coarse to fine detail, it compares at each frequency the signal (the power above the noise floor) with the noise floor. The first frequency where this ratio drops below the threshold, and stays below it, is the cutoff. The resolution is one period of that frequency: 1 / cutoff.

    What the values mean:

    • 1 (default) - the cutoff is where signal and noise are equally strong.
    • 0.5 - accepts detail with half as much signal as noise; finer resolution numbers.
    • 2 - requires twice as much signal as noise; coarser resolution numbers.

    The threshold is applied frequency by frequency; it is not the SNR, dB of the whole area. Keep the same value for all images that are compared.

    Example: three thresholds on the same image

    The cutoffs of thresholds 0.3, 1 and 4 give 8.9, 10.6 and 13.2 nm:

    SNR threshold

  • Number of directions - in how many directions between 0 and 180 degrees the FFT resolution is measured for the astigmatism, up to 72. More directions sample the angle more finely, but each direction rests on fewer data and its value is noisier; the default 36 measures every 5 degrees.

    Example: 12 and 72 directions on an astigmatic image

    Both settings find the same astigmatism:

    Number of directions

  • Smoothing sigma, px - light blurring before the sharpness scores, which reduces the effect of noise on them; 0 turns it off. It does not affect the resolution values. Compare scores only when they were calculated with the same value.

    Example: the same area with sigma 0, 1 and 2

    Smoothing sigma

  • Suppress scan lines - keep it on for scanned images (SEM, FIB-SEM, STEM, confocal); line-to-line intensity jitter would otherwise look like astigmatism along the scan direction.

    Example: a sharp image with scan-line noise

    The orange ellipse is the resolution limit in each direction; with the setting off the stripes stretch it along the vertical axis:

    Suppress scan lines

  • Slow detector - tick it when the detector is slow for the scan speed, so each pixel still holds part of the previous one and the noise is smeared along the scan lines. Without it such images get a far too fine FFT res., a too high SNR, dB and a false astigmatism along the scan direction. MIB recognises these images and asks to tick the box; the scan lines have to run horizontally.

    Example: the focused test slice recorded with a slow detector

    With the setting off the noise floor (dotted) lies too low and the cutoff moves far to the right; with it on, the result is that of the same image from a fast detector:

    Slow detector

Buttons

Buttons panel

  • Clear - removes all rows from the results table, after confirmation.
  • Save - saves the results table, with all plots, to a .mibSharp file.
  • Load - opens a .mibSharp file. When the table already holds measurements, the loaded ones are added after them or replace them.

Report

  • Export - tick any combination of outputs; the files share one base name, each with its own extension:

    • Excel table (.xlsx), CSV table (.csv) - the results table as shown.
    • MATLAB file with all curves (.mat) - the table and all spectra and curves.
    • Report, Word (.docx), Report, PDF (.pdf) - a report in A4 portrait: a summary table first, then one page per measurement with the analysed area and its spectrum, the radial profile, the resolution per direction, a description of the area and the measured values. Writing it takes a few seconds per measurement.
    • Export to MATLAB workspace - the same as the MATLAB file, as a variable with the name given below the checkboxes; only in MIB for MATLAB.

    The export can be cancelled; files written before that are kept.

  • ? - opens this page.
  • Close - closes the window; the results table and the settings are back when it is opened again in the same MIB session.

Reading the results

Column Meaning
Dataset file name of the dataset
Region the analysed area: Shown area, Whole slice, ROI with its label, Selection or Mask with the object number, or Rect with the pixel range of the drawn rectangle
Slice slice number in the shown orientation
Ch color channel
Size, px width x height of the analysed area
Pix. size pixel size, x x y when they differ
FFT res. resolution from the power spectrum
Decorr. res. resolution from the decorrelation analysis
Best res., Worst res. resolution in the sharpest and in the most blurred direction
Astig. ratio worst divided by best resolution; 1 means no astigmatism
Astig. angle most blurred direction in degrees, counterclockwise from horizontal
Grad. energy, Lapl. var. focus scores without units, higher is sharper
Noise SD noise in grey values
SNR, dB signal-to-noise ratio of the area in decibels; -Inf when no signal stands out from the noise
Units unit of the resolution and pixel size values: nm when the dataset has its pixel size in length units
Note FFT: pixel-limited - detail reaches the pixel size, the true resolution may be finer than the 2 pixels reported; record with smaller pixels to measure it. some directions pixel-limited - the same for some directions, so the astigmatism ratio may be too low. FFT: slow detector, not valid - the noise is smeared along the scan lines; tick Slow detector and calculate again. denoised image: resolutions too fine - almost no noise is left, so both resolution values come out far too fine; analyse the image as it comes from the microscope. Decorr.: no peak - the decorrelation analysis found no resolution; use the FFT value

A column stays empty when its method was switched off. The table keeps the measurements of all datasets for comparison; the rows of the active dataset are highlighted. Drag the bar above the table to change its height; dragging it far up hides the Settings panel to make more room, and dragging it back down shows the panel again.

Context menu

Right-click the table for:

  • Jump to area - shows the area of the selected row in the image view, on its slice and time point; the zoom is kept. The dataset the row was measured on has to be the active one, in the same orientation.
  • Remove measurement - removes the selected rows from the table.
warning
  • The resolution values depend on the noise as well as on the sharpness: compare images with similar SNR, dB, and keep the settings the same between the compared rows.
  • SNR, dB and the resolution values assume that the noise of neighbouring pixels is independent, or smeared along the scan lines only by a slow detector (Slow detector). Images that were filtered, denoised, compressed, resampled or averaged along the scan lines show a too high SNR and a too fine resolution; analyse the images as they come from the microscope.

Plot tabs

Click a row of the results table to show it in the tabs. The examples below are slice 3 of the test image, an astigmatic image with the most blurred direction at 30 degrees.

  • Image - the analysed area, with the contrast of the image view at the time of the calculation. When the FFT method was used, an orange line through the centre marks the most blurred direction (Astig. angle); it runs along the streaks of an astigmatic image. The length of the line has no meaning.

    Example: Image tab

    Image tab

  • Spectrum - the power spectrum of the area, with the same orientation as the image. The centre holds the coarse structures, the edge the finest detail the pixel size allows; the brighter a point, the more of that detail the image contains. Where the bright area ends, the signal ends and only noise is left. The frequency axes are in 1/nm: a limit at 0.1 1/nm means 10 nm resolution.

    • Cutoff (dashed blue circle) - the overall resolution limit, FFT res.
    • Directional cutoff (orange curve) - the limit in each direction. A round curve means no astigmatism; an ellipse means astigmatism, and its short axis points along the most blurred direction.
    Example: Spectrum tab

    The orange ellipse is short along 30 degrees, the most blurred direction:

    Spectrum tab

  • Radial profile - the frequency content of the area as curves. The horizontal axis runs from coarse structures on the left to the finest possible detail on the right.

    • Power (blue, left axis) - how much of each detail size the image contains. It falls towards finer detail until it meets the Noise floor (grey dotted), the level of pure noise; the legend gives the noise SD and the SNR of the area.
    • FFT cutoff (orange-red dashed) - where the power is no longer above the noise floor by the SNR threshold; its position gives FFT res.
    • Decorrelation (green, right axis) - the curves of the decorrelation analysis. The solid curve is calculated from the original image, the pale ones from versions with the coarse structures removed. A curve rises while it collects signal and falls where noise takes over; its peak marks where the signal ends.
    • Decorr. cutoff (purple dash-dot) - the highest-frequency peak of all decorrelation curves; its position gives Decorr. res.

    Both vertical lines estimate the same limit and normally lie close together. When they lie far apart, the shape of the blue curve tells which method to trust:

    • Decorr. cutoff far left of FFT cutoff, and the blue curve falls evenly into the noise floor without a bend - the decorrelation analysis fails. This is typical of images whose pixels are coarser than the real resolution; it found a false peak at coarse detail. Use FFT res.
    • FFT cutoff far right of Decorr. cutoff, often at the right edge (pixel-limited), and the blue curve keeps falling at the finest detail instead of running flat along the noise floor - the noise is not independent between pixels. With FFT: slow detector, not valid in Note, tick Slow detector. Otherwise the image was filtered, compressed or resampled: the FFT method then reports a far too fine resolution, and the decorrelation analysis is not reliable either. Analyse an unprocessed image.
    Example: Radial profile tab

    Radial profile tab

  • Directions - the resolution in each direction (blue dots; open circles where the detail reaches the pixel size) and the fitted curve (orange). A flat curve means no astigmatism. With astigmatism the curve forms a wave: its maximum is at the most blurred direction (Astig. angle, Worst res.), its minimum 90 degrees away (Best res.), and the ratio of the two is the Astig. ratio.

    Example: Directions tab

    Directions tab

Test image

Example of the test images

assets\misc\sharpness_test_gold_on_carbon.tif in the MIB folder is a simulated gold-on-carbon specimen with 2 nm pixels, recorded five times: 1 - focused, 2 - defocused, 3 and 4 - astigmatic on either side of focus (most blurred at 30 and 120 degrees), 5 - focused with scan-line noise. Open it and compare the slices with Whole slice.